MDCG 2024-10 - Orphan medical devices

Dr Simon Cumiskey
EU flag

With the introduction of the EU Medical Device Regulation (MDR), there has been an increased emphasis on clinical evidence requirements. Manufacturers of medical devices indicated for rare diseases found these new obligations burdensome, potentially impacting patient access to these devices.

Released in June 2024, the MDCG 2024-10 guidance introduces and defines new terms to the MDR regulatory landscape and aims to address these challenges faced by manufacturers of medical devices indicated for use in rare diseases.

In this article, we explore the impact of MDCG 2024-10 on clinical evidence requirements, shedding light on its key considerations and implications.

Orphan device classification

An “orphan device” is defined as a medical device specifically intended to benefit patients in the treatment, diagnosis, or prevention of a disease or condition that affects no more than 12,000 individuals in the European Union per year. To qualify as an orphan device, at least one of the following criteria must be met:

  1. Insufficiency of alternatives: There are no adequate alternative treatment, diagnostic, or preventive options available for the specific disease or condition.
  2. Clinical benefit: The device offers an expected clinical benefit compared to existing alternatives or the current state of the art. This assessment considers both device-specific factors and the patient population.

Manufacturers seeking orphan device classification must provide a justification based on epidemiological data to the Notified Body. Additionally, this information should be summarised in the Clinical Evaluation Report.

Orphan subpopulation and orphan indication

In addition to the classification of orphan devices, the MDCG 2024-10 provides guidance for devices that serve specific patient sub-populations or have multiple indications.

Orphan subpopulation

This guidance applies to devices intended for use in the treatment, diagnosis, or prevention of a disease or condition within a clinically valid patient sub-population. Specifically, the sub-population should not exceed 12,000 individuals in the European Union per year (within a disease or condition with an annual incidence of more than 12,000 in the EU).

Orphan indication

Sometimes, a medical device may have a specific intended purpose or indication for an orphan population (or orphan subpopulation) in addition to another intended purpose or indication in larger patient populations. In such cases, the MDCG 2024-10 outlines clinical evidence requirements specifically for the component of the intended purpose related to the orphan population.

Clinical vs. non-clinical data for orphan devices

When evaluating an orphan device, both clinical and non-clinical data play crucial roles.

Clinical data considerations

  1. Substantial evidence: If non-clinical data provide substantial high-quality evidence supporting the safety and performance of the orphan device, it can reduce the burden of pre-market clinical data requirements. This evidence can help justify CE marking, even when there are limitations in clinical data that can be addressed through Post-Market Clinical Follow-up (PMCF) activities.
  2. Acceptable limitations: Limitations in pre-market clinical data are acceptable under certain conditions:
    • Expected clinical benefit: There must be sufficient clinical evidence demonstrating the expected clinical benefit, and the device must perform as intended with an acceptable level of safety.
    • Compliance with GSPRs: The device must meet all relevant General Safety and Performance Requirements (GSPRs).
    • PMCF plan: A well-defined PMCF plan should generate additional clinical data within an appropriate time-frame, fully addressing any remaining limitations.

Clinical data sources

Sources of clinical data must relate to either the subject or equivalent devices, and include:

  • Clinical investigations: Studies involving human subjects and undertaken to assess the safety or performance of a device.
  • Peer-reviewed scientific literature: Relevant research published in reputable journals.
  • PMCF: Data collected during post-market follow-up.

Useful non-clinical data

Non-clinical data can complement clinical evidence. Useful sources include:

  • Laboratory and animal tests: Results from in vitro and animal studies.
  • Computer modelling and simulated testing: Data from software-based models, 3D printed models, and other physical simulations.
  • Ex vivo and cadaveric studies: Insights gained from studies using tissue samples or cadavers.
  • Similar devices: Data from devices similar to the orphan device (when equivalence is not demonstrated).
  • State of the Art information: Understanding the technology landscape.
  • Health datasets: Previously collected patient health information, often used for software validation without exposing patients.

Conclusion

MDCG 2024-10 offers guidance on the classification and clinical evidence requirements for orphan devices. Importantly, manufacturers of orphan devices may use non-clinical data, in combination with a rigorous PMCF plan, to supplement clinical evidence to address pre-market clinical data limitations.

Please contact us if you have any questions about this article or would like to understand how Mantra Systems can help you meet your MDR requirements.

Alternatively, for any general regulatory questions you need a quick answer to, have you tried our AI-powered MDR & MDCG chatbot? — it’s free and will answer your queries around the clock.

Related articles

  1. US and EU flags on poles alongside each other.

    Clinical Evidence under EU MDR: Leveraging FDA Clinical Data to Streamline EU MDR Compliance

    FDA approval alone is not sufficient for European market access - a theme we explore futher in this article and the accompanying webinar.

    Chandini Valiya Kizhakkeveetil Chandini Valiya Kizhakkeveetil Regulatory Medical Writer
  2. An AI-generated image of 3 people in an office in front of a whiteboard with the words 'Medical Device Market Entry Strategy' written above a world map.

    EU MDR & NHS DTAC Cybersecurity Requirements for UK Market Entry

    This guest article from our partner Cyber Alchemy shows you how to build cybersecurity evidence for the EU MDR and NHS DTAC.

    Luke Hill Luke Hill Co-Founder of Cyber Alchemy
  3. An illustration showing a GPS-driven navigation route superimposed upon someone using a laptop.

    Where to Launch First? A MedTech Founder's Regulatory Roadmap to the EU, UK and US

    Cyber Alchemy × Mantra Systems — Episode 1: All three markets operate under different regulatory systems and place different demands on manufacturers.

    Ronghe Xu Ronghe Xu Regulatory Medical Writer & Strategic BD Lead China
  4. A woman uses an inhaler.

    Navigating EU MDR Article 117: A Practical Guide to Drug-Device Combination Product Submissions

    Implementation of the EU MDR 2017/745 has brought significant changes.

    Chandini Valiya Kizhakkeveetil Chandini Valiya Kizhakkeveetil Regulatory Medical Writer
  5. NHS CEP logo.

    Mantra Systems becomes an official Programme Partner of the NHS Clinical Entrepreneur Programme

    We're proud to join a this initiative supporting healthcare workers and patients to develop innovative solutions for the NHS.

    Paul Hercock Paul Hercock Chief Executive Officer
  6. Collage art showing a pair of binoculars, an analogy for surveillance.

    How EU MDR Post Market Surveillance differs from FDA post-market expectations

    We compare manufacturer-specific post-market obligations across both regulatory systems.

    Dr Gayle Buchel Dr Gayle Buchel Chief Medical Writer
  7. An arrow arcs from the US over to Europe.

    How EU device classification differs from the US - Are you Prepared?

    Did you know an FDA Class II medical device could be immediately considered as a high-risk Class III device under European Union regulations?

    Gabriela Cardoso Gabriela Cardoso Regulatory Medical Writer
  8. A magnifying glass inspecting a number of wooden cubes with question marks upon them laid upon a blue table. The wooden cube under the magnifying glass has a lightbulb painted on it.

    Fixing the MDR and IVDR? The Commission’s Proposed Amendments and What They Mean for Manufacturers

    Exploring the key elements of this proposal.

    Chandini Valiya Kizhakkeveetil Chandini Valiya Kizhakkeveetil Regulatory Medical Writer
  9. Two arms point at a sign and hold a question mark, in an abstract pop-art style.

    Regulatory Reset? The EU’s Proposed Changes to MDR and IVDR Explained

    Changes published in December 2025 aim to streamline EU medical device and in vitro diagnostics. We explain who is impacted and how.

    Dr Gayle Buchel Dr Gayle Buchel Chief Medical Writer
  10. A pair of glasses rests on an eye test chart.

    Did You Know Your Glasses Were a Medical Device? A Regulatory Guide for Manufacturers

    The importance of correct classification and our recommended path to avoid common ophthalmic device 'gotchas'.

    Gabriela Cardoso Gabriela Cardoso Regulatory Medical Writer

More articles

Need help producing compliant CEPs & CERs? We are offering FREE CEPs to 5 qualifying applicants per week

Get your free CEP